Philips IntelliVue MP series monitor screen malfunction maintenance

March 16, 2018

Latest company news about Philips IntelliVue MP series monitor screen malfunction maintenance

1. Check the machine, the interface does not prompt the touch function failure and other information.

2. Clean the interface with clean water first, the fault keep remains.

3. Enter the maintenance mode to correct the touch program, and the fault is still.

4. Disconnect the machine, and plug and touch part of the data cable to correct the fault still in!The touch screen is damaged, and the model MicroTouch P/N: j510.410 is available.

5. Replace the same type of touch screen for correction, and the MP50 monitor will be OK.

 

The philips IntelliVue MP series monitor requires two codes for calibration.Configuration password: 71034, maintenance password: 1345. Uniform password: 4630.